Cookies on this website

We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

BACKGROUND: Critical care glycaemic control protocols commonly have treatment adjustment (target) ranges spanning ≤2 mmol/L. These require precise point-of-care glucose measurement, unaffected by other variables, to avoid measurement errors increasing glycaemic variability and hypoglycaemic episodes (both strongly associated with mortality in critically ill patients). METHODS: A prospective 206 intensive care patient study was carried out. Arterial glucose concentrations were measured in duplicate using three point-of-care instruments (MediSense Precision PCχ, HemoCue DM and Radiometer 700), a central laboratory instrument (Siemens ADVIA), and in whole blood and plasma using the Yellow Springs Instruments 2300 instrument. RESULTS: Coefficients of variation for the MediSense, HemoCue, Radiometer and Siemens instruments were 5.1%, 2.5%, 2.1% and 2.3%, respectively. Compared with the Siemens instrument, the bias (95% limits of agreement) for the MediSense, HemoCue and Radiometer instruments were 0.0 (-1.4 to 1.4), 0.0 (-1.2 to 1.1) and -0.2 (-0.9 to 0.6) mmol/L, respectively. The whole blood-plasma glucose concentration difference was significantly affected by the haematocrit. MediSense and HemoCue instrument performances were substantially affected by haematocrit. MediSense instrument performance was also affected by pH and PaO(2). Radiometer instrument performance was not affected by haematocrit, pH or PaO(2). CONCLUSIONS: The MediSense instrument was too imprecise for use in critically ill patients. The haematocrit range seen is too great to allow fixed-factor conversion between whole blood and plasma values, substantially affecting the accuracy of both glucose meters. However, the Radiometer instrument was unaffected by the haematocrit, pH or pO(2), resulting in a performance equivalent to the laboratory method. Instrument performance differences may therefore partially explain the differing results of tight glycaemic control therapy trials.

Original publication

DOI

10.1258/acb.2011.011162

Type

Journal article

Journal

Ann Clin Biochem

Publication Date

03/2012

Volume

49

Pages

144 - 151

Keywords

Blood Glucose, Critical Illness, Female, Hematocrit, Humans, Hydrogen-Ion Concentration, Male, Middle Aged, Oxygen, Prospective Studies, Reproducibility of Results