We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

PURPOSE: Ghosting-robust reconstruction of blipped-CAIPI echo planar imaging simultaneous multislice data with low computational load. METHODS: To date, Slice-GRAPPA, with "odd-even" kernels that improve ghosting performance, has been the framework of choice for such reconstructions due to its predecessor SENSE-GRAPPA being deemed unsuitable for blipped-CAIPI data. Modifications to SENSE-GRAPPA are used to restore CAIPI compatibility and to make it robust against ghosting. Two implementations are tested, one where slices and in-plane unaliasing are dealt in the same serial manner as in Slice-GRAPPA [referred to as one-dimensional (1D)-NGC-SENSE-GRAPPA, where NGC stands for Nyquist Ghost Corrected] and one where both are unaliased in a single step (2D-NGC-SENSE-GRAPPA), which is analytically and experimentally shown to be computationally cheaper. RESULTS: The 1D-NGC-SENSE-GRAPPA and odd-even Slice-GRAPPA perform identically, whereas 2D-NGC-SENSE-GRAPPA shows reduced error propagation, less residual ghosting when reliable reference data were available. When the latter was not the case, error propagation was increased. CONCLUSION: Unlike Slice-GRAPPA, SENSE-GRAPPA operates fully within the GRAPPA framework, for which improved reconstructions (e.g., iterative, nonlinear) have been developed over the past decade. It could, therefore, bring benefit to the reconstruction of SMS data as an attractive alternative to Slice-GRAPPA. Magn Reson Med 77:998-1009, 2017. © 2016 The Authors Magnetic Resonance in Medicine published by Wiley Periodicals, Inc. on behalf of International Society for Magnetic Resonance in Medicine. This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.

Original publication

DOI

10.1002/mrm.26179

Type

Journal article

Journal

Magn Reson Med

Publication Date

03/2017

Volume

77

Pages

998 - 1009

Keywords

CAIPI, GRAPPA, Nyquist ghost, SENSE, multiband, simultaneous multislice, Algorithms, Artifacts, Echo-Planar Imaging, Image Enhancement, Image Interpretation, Computer-Assisted, Motion, Phantoms, Imaging, Reproducibility of Results, Sensitivity and Specificity, Signal Processing, Computer-Assisted