We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

The aim of this work was to compare the performances of the Projector Operator (PO) and the independent component analysis (ICA) based method for fetal signal extraction from fMCG. Both methods were applied on data recorded with the SARA system installed at the UAMS, Little Rock (Arkansas-USA), and with the ATB system installed at ITAB-Chieti (Italy). Twenty datasets recorded with both systems were analysed. The detection rates, signal-to-noise ratios, fetal signal amplitude reconstructions and residual maternal contamination of the reconstructed fetal signals were calculated, and the PO and ICA-based approach provided fetal signals with similar features, although ICA seemed to be more effective when noisy data were processed. © 2007 Elsevier B.V. All rights reserved.

Original publication

DOI

10.1016/j.ics.2007.01.002

Type

Journal article

Journal

International Congress Series

Publication Date

01/06/2007

Volume

1300

Pages

733 - 736