We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

BACKGROUND: Bedside lung volume measurement could personalise ventilation and reduce driving pressure in patients with acute respiratory distress syndrome (ARDS). We investigated a modified gas-dilution method, the inspired sinewave technique (IST), to measure the effective lung volume (ELV) in pigs with uninjured lungs and in an ARDS model. METHODS: Anaesthetised mechanically ventilated pigs were studied before and after surfactant depletion by saline lavage. Changes in PEEP were used to change ELV. Paired measurements of absolute ELV were taken with IST (ELVIST) and compared with gold-standard measures (sulphur hexafluoride wash in/washout [ELVSF6] and computed tomography (CT) [ELVCT]). Measured volumes were used to calculate changes in ELV (ΔELV) between PEEP levels for each method (ΔELVIST, ΔELVSF6, and ΔELVCT). RESULTS: The coefficient of variation was <5% for repeated ELVIST measurements (n=13 pigs). There was a strong linear relationship between ELVIST and ELVSF6 in uninjured lungs (r2=0.97), and with both ELVSF6 and ELVCT in the ARDS model (r2=0.87 and 0.92, respectively). ELVIST had a mean bias of -12 to 13% (95% limits=±17 - 25%) compared with ELVSF6 and ELVCT. ΔELVIST was concordant with ΔELVSF6 and ΔELVCT in 98-100% of measurements, and had a mean bias of -73 to -77 ml (95% limits=±128 - 186 ml) compared with ΔELVSF6 and -1 ml (95% limits ±333 ml) compared with ΔELVCT. CONCLUSIONS: IST provides a repeatable measure of absolute ELV and shows minimal bias when tracking PEEP-induced changes in lung volume compared with CT in a saline-lavage model of ARDS.

Original publication

DOI

10.1016/j.bja.2019.11.030

Type

Journal article

Journal

Br J Anaesth

Publication Date

03/2020

Volume

124

Pages

345 - 353

Keywords

ARDS, computed tomography, inspired sinewave technique, lung volume measurement, method comparison, Animals, Disease Models, Animal, Lung Volume Measurements, Point-of-Care Testing, Positive-Pressure Respiration, Reproducibility of Results, Respiratory Distress Syndrome, Adult, Saline Solution, Sus scrofa, Tomography, X-Ray Computed