We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

Cardiac index was measured using thoracic bioimpedance (CIbi) and thermodilution (CItd) in 19 patients with proven sepsis, undergoing artificial ventilation of the lungs. There was a poor correlation between the techniques (r = 0.36, 242 data sets, regression line CIbi = 0.16 CItd + 2.56 litre min-1 m-2). The overall bias (CItd-CIbi) was 1.69 litre min-1 m-2 with limits of agreement (precision) of +4.17 to -0.79 litre min-1 m-2. In individual patients the bias was from -0.46 to 4.56 litre min-1 m-2 with the limits of agreement from +/- 0.29 to +/- 2.55 litre min-1 m-2 around the bias values. The two techniques cannot be used interchangeably in this group of patients.

Type

Journal article

Journal

Br J Anaesth

Publication Date

01/1993

Volume

70

Pages

58 - 62

Keywords

Bacterial Infections, Body Surface Area, Cardiac Output, Electric Impedance, Humans, Respiration, Artificial, Thermodilution, Thorax